Disinfecting the Machine
From the viewpoint of infectious disease prevention measures, we will explain how to disinfect the parts of the product that come into contact with the hands.
Read This before Cleaning and Disinfecting the Product
Be sure to turn off the power of the product and disinfect it with the power plug and battery removed.
When cleaning the product, spray the disinfectant lightly on a soft, dry cloth, and wipe the area lightly and quickly where the product come into contact with the hands.
Do not spray the disinfectant directly on the product (because the product is a precision instrument, it may malfunction or smoke due to corrosion, and so on if chemicals enter the inside of it).
Be sure not to allow water to get into the ventilation openings, the gaps around the operation buttons, and the connectors and battery terminals.
Prohibited solvents
Do not use chemicals such as cleaners and organic solvents such as flammable sprays, thinner, or benzine. They may seep through the gap or melt the plastic of the main body, causing a malfunction or fire.
Do not use detergents containing bleach or abrasives.
Precautions after disinfecting floors
After disinfecting floors and walls, wipe disinfectants off with a dry cloth so that no disinfectants pool in the space near our products (floors, walls, and so on). The intake fan of the product may suck moisture into the machine.